Endress+Hauser iTHERM ModuLine TM151 industrial thermometer assembly, with thermowell and insertion details to confirm before RFQ.
Temperature measurement products support process monitoring, quality control and equipment protection in industrial applications.
Endress+Hauser iTHERM ModuLine TM151 RFQs should confirm the thermometer assembly, process connection and insertion details before replacement. Endress+Hauser technical information places TM151 within modular industrial thermometers, so the thermowell, sensor element and installation conditions should be checked before selecting a matching assembly.
TM151 replacement: distinguish U, L and IL
Do not use the thermowell length as the replacement measuring-insert length. In the TM151 technical information, TI01707T/EN/04.25-00, U is immersion length, L is thermowell length (U + T), and IL is measuring-insert length; T identifies the thermowell lagging length.
For a replacement, match the installed assembly’s order code and dimensional drawing. Record the thermowell bore, neck arrangement, sensor connection and approval version, then use the corresponding IL calculation. Neck and head dimensions, bottom thickness, gasket allowance and spring travel depend on the configuration. A formula from a different neck or protection variant is not a reliable substitute.
For sourcing, send order code, stem or thermowell details, insertion length, process connection, sensor type, transmitter option, material notes, quantity and nameplate photos. Related entry pages: Endress+Hauser, Temperature Measurement, Power Generation and RFQ.
For sourcing, Shengshi treats Endress+Hauser iTHERM ModuLine TM151 Industrial Thermometer as a configured industrial product. Buyers should confirm the exact model code, quantity, application conditions and required documentation before quotation.
Measurement categoryTemperature Measurement
BrandEndress+Hauser
ModeliTHERM ModuLine TM151
IndustriesChemical, Oil & Gas, Power Generation